| Weight | 
			0.290 kg | 
		
			
			| Dimensions | 
			18 × 13 × 6 cm | 
		
			
			| Communication | 
			BLE – Bluetooth Low Energy. 
 | 
		
			
			| Ketone warning | 
			Yes 
 | 
		
			
			| Power source | 
			2 x AAA Batteries 
 | 
		
			
			| Memory capacity | 
			1000 sets (meter) / Unlimited sets (connected to ZAF-552) 
 | 
		
			
			| Day average (Glucose) | 
			7, 14, 21, 28, 60 and 90 days 
 | 
		
			
			| Daily alarms | 
			4 daily alarms 
 | 
		
			
			| Dimensions | 
			10,2 (L) x 5,7 (W) x 2,2 (H) cm 
 | 
		
			
			| Weight | 
			57 g. without battery 
 | 
		
			
			| Operating conditions | 
			Temperature: +8°C to +45°C 
Humidity: 10% to 90% R.H. 
 | 
		
			
			| Storage contitions | 
			Temperature: -20°C to +60°C (meter), +2°C to +30°C (Strips) 
 | 
		
			
			| Glucose test strip | 
			Product: ZAF-143-2 
Enzyme type: GDH-FAD 
Sample size: 0,5uL 
Reaction time: 5 sec. 
Measurement range: 10-800 mg/dl (0,56-44,4 mmol/L) 
Hematocrit range: 0% – 70% 
Precision: SD <5mg/dl (0,278 mmol/L), CV <5% at 100mg/dl (5,55 mmol/L) 
Accuracy: +/- 15 mg/dl if < 100 mg/dl, +/-15% if >100 mg/dl 
 | 
		
			
			| Ketone test strip | 
			Product: ZAF-143-3 
Enzyme type: HBD 
Sample size: 0,8uL 
Reaction time: 10 sec. 
Measurement range: 0,1 – 8,0 mmol/L 
Hematocrit range: 10% – 70% 
Precision: < 1 mmol/L SD < 0,1 mM, > 1 mmol/L, CV < 7,5% 
 | 
		
			
			| Lactat test strip | 
			Product: ZAF-143-5 
Sample size: 0,8uL 
Reaction time: 5 sec. 
Measurement range: 0,3 – 22 mmol/L 
Hematocrit range: 10% – 65% 
Precision: ≤ 3 mmol/L SD < 0,3 mM, > 3 mmol/L, CV < 7,5% 
 | 
		
			
			| Total cholesterol test strip | 
			Product: ZAF-143-4 
Sample size: 3,0uL 
Reaction time: 60 sec. 
Measurement range: 100 – 400 mg/dl 
Hematocrit range: 20% – 60% 
Precision: CV < 7,5% 
 | 
		
			
			| Uric acid test strip | 
			Product: ZAF-143-6 
Sample size: 0,5uL 
Reaction time: 15 sec. 
Measurement range: 3 – 20 mg/dl (178 – 1190 μmol/L) 
Hematocrit range: 20% – 60% 
Precision: CV ≤ 5 mg/dl, SD < 0,5 mg/dl; > 5mg/dl, CV < 7,5% (≤297 μM, SD 297 μM, CV < 7,5%) (≤0,29mM, SD0,29mM, CV<7,5%) 
 | 
		
			
			| Triglycerides test strip | 
			Product: ZAF-143-19 
Sample size: 3,0uL 
Reaction time: 60 sec. 
Measurement range: 70 – 600 mg/dl 
Hematocrit range: 20% – 55% 
Precision: 70-125mg/dl. SD<10mg/dl : 126-600 mg/dl CV < 7,5% 
 |